DOUBLE-LUMEN CYSTOMETRY CATHETER

You are here:

FEATURES:

  • Allow catheterization of the patient with simultaneous bladder filling and pressure measurement

  • Feature asymmetrical lumens, with the filling lumen larger than the pressure measurement lumen

  • Designed so that only 30% of the cross-sectional area is dedicated to bladder pressure measurement, while 70% is dedicated to filling, this enables effective bladder filling at an appropriate rate using a smaller-diameter catheter. For example, to achieve a maximum filling rate over 100 ml/min, an 8Ch or 10Ch catheter is sufficient

  • Enables more accurate pressure measurement during voiding

  • Provides greater ease and comfort for the patient

  • Includes extension lines of sufficient length, reaching from the patient to the urodynamic device

  • Eliminates the need to purchase separate extension lines
  • Reduces the number of unnecessary connections, each of which is a potential site for micro air bubble formation that can impact pressure measurement accuracy